Where is the drain valve on the Thermalator Heating Units?
The Thermalator Heating Units has an easily accesbile rear drain.
What is the lowest temperature the Thermalator Heating Units will operate?
The water temperature for normal operating conditions for the Thermaltor Heating unit is about 165°F. The thermostat is very sensitive; even a small adjustment can raise or lower the temperature several degrees. Facing the back of the unit, find the metal temperature knob between the two screws on the lower right side of the unit. Using the screwdriver, turn the stem counter-clockwise to decrease the temperature. Allow the temperature to stabilize, and measure the water temperature with the thermometer. Note that a small adjustment changes several degrees.